Accredited with ISO 9001:2000 Certification, Rishabh Instruments has world-class facilities to test instruments as per IEC, DIN, IS and ANSI standards. The company’s products also have UL, CSA, and CSE certification. The QA lab is equipped with highly accurate standards for calibration and testing equipments like Datron Wavetek-1271, Fluke 5500A, Zera (RMM), Yokogawa Current & Voltage standards, Vibration Tester, Environment Chamber High Voltage Tester, Insulation Tester from reputed companies in Japan, Germany and Switzerland. Presently the company is in the process of setting up of a NABL accredited Laboratory.
Multifunction Instruments, Insulation Tester, Energy Meters, Power Transducer.
Quality Assurance System In Rishabh Instruments Pvt. Ltd.
 

The Quality Management System of Rishabh Instruments Pvt. Ltd. is in accordance to ISO 9001:2000, since April 1985.

QAS as a department is subdivided into four parts viz. Inward, In process, Final and Testing & Calibration.

 

Inward:

All incoming material into the factory is inspected at this section. Statistical quality control technique used is VQR (Vendor Quality Rating) and the sampling procedure used is as per IS 2500. Two Engineers and two operators are stationed in this section to cater to Electronic and mechanical component requirement.

In process:

The basic concept of this section is to built in “quality” in the processes itself. Critical stations are identified and setting patrol and final inspections are done at these stations. For special processes operator qualification and operator training is done. In process Quality plans are available to guide through the process of In process inspections

Final:

Final products are inspected at this section. The inspection is in two parts viz. electrical and packing. Quality plans are exclusively made per product to guide through the process of final inspection. The sampling procedure made is as per IS 1248. The statistical quality control technique used is D Merit Index

Testing and Calibration:

Rishabh Testing and Calibration laboratory equipped with modern amenities to carry out the followings:

1. Type testing on different final products (for verification of design) as per IS/IEC/DIN/EN standards.
2. Approval of components.
3. In house calibration programme of all the instruments/ sources/ standards held by the factory.
4. To co-ordinate the external calibration of standards.

Rishabh Testing and calibration Laboratory is going for NABL accreditation (National Accreditation board for Testing and Calibration Laboratories) as per ISO/IEC 17025:2005, for its calibration facilities in the Electro-technical field.

Testing Setup:

1. Environmental Testing Temperature cum Humidity Chamber
Make: Espec Corporation, Japan
Temperature Range: -40 to 150 Deg C, Accuracy: ±0.3˚C Humidity Range: 20 to 98 %RH,
Accuracy: ±2.5%RH
2. Stresses owing to vibrations Vibration Tester with Precision Piezoelectric sensor.
Make: Saraswati Dynamics, Noida
Max payload capacity: 30kg
Max acceleration: 80g
Max displacement: 15mm p-p
Frequency range: 5kHz
Modes: Sweep mode control, discrete step control.
3. 1Φ/3Φ Power Reference Multi Meter and Source for Multi-Transducer, Power Transducer & PF Transducer, Multi-Function Instruments, Watt & VAr meters
 
a. RMM 3001, Make: Zera, Germany, Accuracy: ±0.02%
b. VCS-320, Make: Zera, Germany, Accuracy: ±0.2%
c. DLP-03, Make: Eberle, Accuracy: ±0.1%
4. 1Φ/3Φ Energy Meters
Reference Standard RS2310A with Power Unit 4000A
Make: Applied Meters, Accuracy: ±0.05%
5. Synthesized function generator for Frequency Transducer
Yokogawa FG120, Accuracy: ±20ppm
6. Multi Function Calibrators for Digital Multimeter
 
a. Fluke 5500A, Accuracy: ±50ppm
b. Fluke 5520A, Accuracy: ±12ppm
c. Wavetek 9100, Accuracy: ±60ppm
7. Voltage –Current Standard for Analog& Digital Panel Meters.
 
a. Yokogawa 2558 AC Standard, Accuracy ±0.08%
b. Yokogawa 2560 DC Standard, Accuracy ±0.02%
8. High Voltage Test
HV Breakdown Tester
Make: Laxtronics
9. Insulation Resistance Tester
Rish Insu 5000A
Make: Rishabh
10. Test for Temperature Rise
Temperature Probe with Digital Display,
Make: BBC Microwatts

Calibration Facilities:

1. Active & Reactive AC Power (1Ǿ&3Ǿ), Phase Angle 0.1lag-1-0.1lead
Standard Equipment: RMM 3001,
Make: Zera, Germany, Accuracy: ±0.02%
2. AC-DC Voltage –Current
Standard Equipment: 8 ½ Digits DMM,
Make: Wavetek Datron, Accuracy: ±10ppm
3. Hi-Meg Resistance
Standard Equipment: 8 ½ Digits DMM,
Make: Wavetek Datron, Accuracy: ±10ppm
4. 4. Frequency
Standard Equipment: Metrahit 28S
Make: Gossen, Accuracy: ±0.05%
5. 5. Digital Multimeter
Standard Equipment: Multi-Function Calibrators Fluke 5500A
Make: Fluke, Accuracy: ±50ppm

Retention periods of certificates:

For all the instruments with which test certificates are provided by Rishabh Instruments Pvt. Ltd. the retention period is Three Years. The soft copy of the test certificates of DMM is available on the centralized computer server system whereas the photocopy of all other hard copies is available with QAS.

International Approvals for our Products:

1. CSA approval for Digital Multimeter & Transducer,
2. UL approval for Analog Panel, relays and
3. CE Declaration for Analog Panel Meter, DMM, Transducer
Multifunction Instruments, Insulation Tester, Energy Meters, Power Transducer.